Covidien/Valleylab Force Triad Electrosurgical Generator

SKU: EN-VL-FORCETRIAD
Condition
Category
LEAD TIME: Estimated Ship Time: 7-10 Days
PACKAGING: 1
WARRANTY: 12 Months
SHIPPING/RETURNS:
This product can be shipped to any location within the continental United States using FedEx Ground, UPS Ground, or Freight services. Shipping rates will be determined based on the destination zip code. For further information, please refer to our Shipping and Return policies.

Listing Terms & Conditions

All product and company names are trademarks of their respective holders. Use of them does not imply any affiliation with or endorsement or sponsorship by them.

The Covidien/Valleylab Force Triad energy platform is a comprehensive electrosurgical system that offers a wide range of surgical capabilities including precise cutting and coagulation, bipolar functionality, and efficient vessel sealing. This advanced generator combines all these features into a single unit, making it suitable for both open and laparoscopic procedures. The system incorporates three cutting-edge technologies – TissueFect, Valleylab, and LigaSure fusion – to enhance surgical precision and efficiency. With its intuitive interface consisting of three touchscreens, users can easily control and manage the functions of the system.

FEATURES:

  • TissueFect sensing technology to precisely manage energy delivery.
  • Enhanced LigaSure tissue fusion technology that fuses vessels up to 7mm, lymphatics, pulmonary vasculature, and tissue bundles.
  • alleyLab mode that provides a unique combination of monopolar hemostasis and dissection.
  • Automatic instrument recognition.
  • Compatible with conventional electrosurgical instruments and all current and new LigaSure instruments.

    SPECIFICATIONS:

    • Manufacturer: Covidien/Valleylab
    • Part Number: (OEM): VALLEYLAB FORCE TRIAD
    • Category: Electrosurgical Units
    • Condition: Refurbished
    • Weight: 13.6lbs.
    • Warranty: 3 Months

    • General:
      • Output Configuration: Isolation output.
      • Cooling: Natural convection and fan.
      • Display: Three LCD touchscreens.
      • Connector ports: LED illuminated Smart connector readers
        • ForceTriad energy platform cart (FT900) Universal.
        • Mounting cart (UC8009), and/or the UC8010 Over-shelf.
        • Operating room boom systems.
        • Any Stable, flat surface such as a table or cart top.

    • Equipment Dimensions and Weight:
      • Height: 25.5 cm (10 in.)
      • Width: 45.8 cm (18 in.)
      • Depth: 50.8 cm (20 in.)
      • Weight: 13.6 kg (30 lb)

    • Activation Tone:
      • The audio levels stated below are for activation tones (cut, Valleylab, coag, bipolar, and LigaSure Modes) and alarm tones (REM and system alarms) at a distance of one meter.

    • Volume and Frequency:
      • Volume (adjustable): 45 to 65 dBA
      • Frequency Cut: -660 Hz
        • Valleylab: 800 Hz.
        • Coag: 940 Hz.
        • Bipolar: 940 Hz.
        • LigaSure: 440 Hz.
      • Duration: Continuous while the system is activated.

    • Alarm Tone:
      • Volume (adjustable):>65 dBA.
      • Frequency: -REM: 660 Hz.
        • Regrassp: Two tones: High= 985 Hz, Low = 780 Hz.
        • Seal Complete: 985 Hz.
        • Error/System Alert: Beep tone = 1400 Hz.
        • Duration:
          • REM: Two 1/2 second tones separated by 1/2 second for each REM event
          • Reactivate/Regrassp: Four 175 msec tones-high, low, high– low separated by 1/2 second.
          • Check Instrument: Six 175 msec tones – high, low, high low, high, low -Seal Complete: Two 175 msec tones separated by 175 msec for each Seal Complete event.
          • Error/System Alert: Two 1/2 second tones separated by 1/2 second for each Error/ System Alert event.

    • REM Contact Quality Monitor:
      • Interrogation frequency:80 kHz ± 10 kHz
      • Interrogation current: < 100μA
      • Interrogation voltage: < 12V RMS

    • Acceptable Resistance Range:
      • REM resistance measurements are ± 10% during RF activation and ± 5% when RF output is not activated.
      • REM patient return electrode: 5 to 135 ohms or up to a 40% increase in the initial measured contact resistance (whichever is less).
      • If the measured resistance is outside the acceptable range(s) noted above, a REM fault condition occurs.

    • REM Alarm Activation:
      • REM patient return electrode: When the measured resistance exceeds the standard range of safe resistance (below 5 ohms or above 135 ohms) or when the initially measured contact resistance increases by 40% (whichever is less), the REM alarm indicator enlarges and flashes red and yellow, a tone sounds twice, and RF output is disabled. The indicator remains illuminated red and yellow until you correct the condition causing the alarm. Then, the indicator illuminates green, and RF output is enabled.

    • Autobipolar:
      • The ForceTriad energy platform is equipped with an auto bipolar feature that allows for the automatic activation of bipolar energy.
      • Note: The autobipolar electrode function requires the use of the Valleylab E0018 bipolar Instrument cord.
      • Autobipolar Specifications:
        • Interrogation frequency: 80 kHz ± 10 kHz.
        • Interrogation current: < 100 μA.
        • Activation impedance: 20 Ω to 500.
        • Deactivation impedance: User selectable: 1,500 Ω 1,800 Ω or 2,200 Ω.
        • Measurement accuracy: ± 14% of Full-Scale activation impedance while keying active above 10 watts.
        • + 50%, -14% of full-scale activation impedance while keying active at 10 watts or below.
        • ± 5% of Full-Scale activation impedance while keying inactive.
        • Keying delay: User selectable in 500 msec increments from 0 sec to 2.5 sec

    • Duty Cycle:
      • Under maximum power settings and rated load conditions, the ForceTriad energy platform is capable of operating a duty cycle of 25% defined as 10 seconds active and 30 seconds inactive, in any mode for 4 hours.
      • Caution
        • Use of duty cycles greater than 25% (10 seconds active followed by 30 seconds inactive) will increase the risk that heat build-up under a return electrode may be high enough to injure the patient. Do not continuously activate for longer than one minute.

    • Low Frequency (50/60 Hz) Leakage Current:
      • Enclosure source current, ground open <300 μA.
      • Source current, patient leads, all outputs
        • Normal polarity, intact ground: < 10 μA.
        • Normal polarity, ground open: < 50 μA.
        • Reverse polarity, ground open: < 50 μA.
        • Mains voltage on applied part: < 50 μA.
      • Sink current at high line, all inputs<50 μA

    • Input Power:
      • 100-120 Volt
        • Maximum VA at nominal line voltage:
          • Idle: 52 VA
          • Bipolar: 450 VA
          • Cut: 924 VA
          • Coag: 530 VA
        • Input Mains Voltage, Full Registration Range:
          • 90-132 Vac
        • Input Mains Voltage, Operating Range:
          • 85-132 Vac

      • 220-240 Volt:
        • Maximum VA at nominal line voltage:
          • Idle: 52 VA
          • Bipolar: 450 VA
          • Cut: 924 VA
          • Coag: 530 VA
        • Input Mains Voltage, Full Registration Range:
          • 208-264 Vac
        • Input Mains Voltage, Operating Range:
          • 170-264 Vac

      • 100-120 Volt:
        • Mains Current (Maximum):
          • Idle: 0.4 A
          • Bipolar: 2.0 A
          • Cut: 7.0 A
          • Coag: 4.0 VA
          • LigaSure: 5.0 VA
        • Mains Line Frequency Range (Nominal):
          • 50 Hz to 60 Hz
        • Fuses (2): 5mm x 20 mm A8, 250 V:
          • Fast Blow
        • Power cord: 3-prong hospital grade:
          • Connector

      • 220-240 Volt:
        • Mains Current (Maximum):
          • Idle: 0.2 A
          • Bipolar: 1.0 A
          • Cut: 3.5 A
          • Coag: 2.0 VA
          • LigaSure: 2.5 VA
        • Mains Line Frequency Range (Nominal):
          • 50 Hz to 60 Hz
        • Fuses (2): 5mm x 20mm A8, 250 V:
          • Fast Blow
        • Power cord: 3-prong hospital grade:
          • Connector

                    In addition please acknowledge the following disclaimer:

                    * IMPORTANT - PLEASE READ BEFORE PURCHASING! * All pictures of items being sold are used for reference purposes and are described as accurately as possible. If there is an error in a listing, we will try and resolve any issues to the best of our abilities that may arise before or after a sale. Complete units all come with a power cord and DO NOT come with any cables, sensors, manuals, accessories etc. unless otherwise noted in the description.

                    Purchaser agrees that seller is not liable for any injuries or damages of any kind associated in any way with this equipment. Purchaser agrees to indemnify and hold harmless seller and seller's officers, directors, employees, agents, and shareholders against any and all claims, liabilities, obligations, demands, cost and expenses of every kind and nature, including but not limited to attorney's fees and litigation expenses, from or related to the equipment and any use thereof.

                    * NOTE - FDA Regulation and Certification: "The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulation, we will verify your status as an authorized purchaser of this item before shipping of the item."