Surgeons can benefit from instant response technology, which enhances their performance and reduces the risk of tissue damage and neuromuscular stimulation. It also eliminates the need to manually increase the power settings. The Force EZ generator, equipped with this technology, automatically adjusts itself to adapt to tissue changes during surgery. This ensures a constant power delivery and minimizes any resistance encountered.
One of the advantages of using instant response technology is the significant reduction in capacitive coupling, which can be up to 30-50%. This reduction is achieved by controlling the RMS voltage and high-frequency harmonics. By lowering the voltage, the technology minimizes neuromuscular stimulation and ensures precise energy delivery, thereby minimizing any collateral damage.
If you are in search of more Valleylab electrosurgical units, you might be interested in the Covidien Valleylab Force FX. This device also incorporates instant response technology, featuring an advanced feedback system. This system can detect tissue changes an impressive 200 times per second, allowing the generator to adjust voltage and current accordingly. This ensures that the power output remains appropriate for the specific surgical requirements.
What sets instant response generators apart is their high power efficiency rating (PER), which is made possible by their smart generator technology. PER indicates how accurately a generator can deliver the selected power across a wide variety of tissue types. With a remarkable PER of 97, the Force EZ generator consistently achieves the desired surgical effect.
In summary, instant response technology revolutionizes the field of surgery by providing surgeons with improved performance and reducing the risk of tissue damage. The Force EZ generator, along with other Valleylab electrosurgical units like the Covidien Valleylab Force FX, exemplifies this technology with their advanced feedback systems and high power efficiency ratings.
FEATURES:
- Instant Response technology ensures that the power delivered remains virtually constant, regardless of the tissue type
- Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
- Three internal microcontrollers reduce system reaction time and increase the system's processing speed
- Easy to use, cost-effective system when performance is required and the price is of equal concern
- Two cut modes, both controlled by Instant Response technology, offer surgeons a variety of choices:
- Pure Cut for a clean, precise cut
- Blend for cutting with hemostatsis
- Two primary coag modes:
- Low for lower voltage contact coagulation suitable in laparoscopic and delicate tissue work
- High for efficient non-contact coagulation in most applications
SPECIFICATIONS:
- Manufacturer: Covidien/Valleylab
- Part Number: (OEM): VALLEYLAB FORCE Ez
- Category: Electrosurgical Units
- Condition: Refurbished
- Weight: 15.0lbs.
- Warranty: 3 Months
- Dimensions:
- Height: 5 inches (12.7 cm)
- Width: 16 inches (40.6 cm)
- Length: 15.6 inches (39.5 cm)
- Weight: < 15lb (6.8 kg)
- Output Waveforms:
- Bipolar: 470 kHz sinusoid
- Monopolar Cut:
- Pure: 393 kHz sinusoid
- Blend: 393 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope
- Monopolar Coag:
- Desiccation:
- Low 1: 240 ± 40 kHz sinusoid recurring at 39 kHz. 8% duty cycle
- Low 2: 393 kHz sinusoid
- Low 3: 393 kHz sinusoid
- Fulguration:
- High 1: 470 ± 40 kHz damped sinusoidal bursts with a repetition frequency of 57 kHz
- High 2: 470 ± 40 kHz damped sinusoidal bursts with a repetition frequency of 30 kHz
- Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 90-135 volts (into a 300-ohm load)
- Low-Frequency Leakage (50-60 Hz):
- Source current, patient leads, all outputs tied together
- Normal polarity, intact chassis ground: < 10µA
- Normal polarity, ground open: < 50µA
- Reverse polarity, ground open: < 50µA
- Sink current, 140V applied, all inputs: < 50µA
- High Frequency Leakage:
- Bipolar: Less than 60 mArms
- Monopolar: Less than 150 mArms
- Operating Parameters:
- Ambient Temperature range: 10° to 40° C (50° to 104° F)
- Relative humidity: 30% to 75%, non-condensing
- Atmospheric Pressure: 700 to 1060millibars
- Warm-up Time: If you transported or stored the generator at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use
- Transport & Storage:
- Ambient Temperature range: 34° to 65° C (-29° to 149° F)
- Relative Humidity: 25% to 85%, non-condensing
- Atmospheric Pressure: 500 to 1060 millibars
- Ambient Temperature: 34° to 65° C (-29° to 149° F)
- Relative Humidity: 25% to 85%
- nNoncondensing Atmospheric Pressure: 500 to 1060 millibars
- Duration of storage: If you store the generator longer than one year, replace the battery for battery-backed RAM and complete a full test and recalibration before use.
- Input Power Requirements:
- 110-120Vac:
- Operating Range: 85-140Vac
- Max Current Cut: 8A
- Max Current Coag: 4.2A
- 220-240Vac:
- Operating Range: 170-280Vac
- Max Current Cut: 4A
- Max Current Coag: 2.1A
In addition please acknowledge the following disclaimer:
* IMPORTANT - PLEASE READ BEFORE PURCHASING! * All pictures of items being sold are used for reference purposes and are described as accurately as possible. If there is an error in a listing, we will try and resolve any issues to the best of our abilities that may arise before or after a sale. Complete units all come with a power cord and DO NOT come with any cables, sensors, manuals, accessories etc. unless otherwise noted in the description.
Purchaser agrees that seller is not liable for any injuries or damages of any kind associated in any way with this equipment. Purchaser agrees to indemnify and hold harmless seller and seller's officers, directors, employees, agents, and shareholders against any and all claims, liabilities, obligations, demands, cost and expenses of every kind and nature, including but not limited to attorney's fees and litigation expenses, from or related to the equipment and any use thereof.
* NOTE - FDA Regulation and Certification: "The sale of this item may be subject to regulation by the U.S. Food and Drug Administration and state and local regulatory agencies. If so, do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulation, we will verify your status as an authorized purchaser of this item before shipping of the item."